The InfraCal 2 Ethanol Blend Analyzer for measuring percent ethanol in gasoline is the ideal solution where a quick, accurate, on-site measurement is required and a single, repetitive ethanol blend analysis is needed.  The InfraCal 2 ATR-E is based on proven IR technology, and used by fuel terminals, laboratories and regulatory agencies worldwide.  The compact, easy-to-use InfraCal 2 ATR-E has a measurement range of 0-20 & 70-98%% ethanol in gasoline and an accuracy of +/- 0.2%. These results are comparable to ASTM D4815.

Samples are applied to a  zinc selenide attenuated total reflectance (ATR) sample plate and easily cleaned with a cloth & isopropyl alcohol.  No solvents requiring special ventilation, expensive waste disposal or difficult to clean cells are needed. The analysis can be done by non-technical personnel with no formal training and is ideal for lab or field use. The InfraCal2 ATR-E has a touch screen display, data storage and transfer, a multi-language interface and password protected settings.  The internal battery pack allows for portability and up to 12 hours continuous use, providing results in under a minute at a fuel loading rack, in a laboratory or at a pump in a fueling station.

  • Measurement range: 0 – 20 & 70 – 98% ethanol in gasoline
  • Instrument repeatability: +/- 0.2%
  • Compares to ASTM D4815

Highlights

  Benefits   Features
Internal Data Storage Data storage including each measurement tagged with analyst, location and date/time information, and password protected instrument settings
Export Directly to your PC via USB or a Flash Drive
Touch Screen Display The touch screen display allows a simple set up for quick operation along with an easy calibration wizard.
Internal Battery The internal battery pack allows for portability and up to 6 hours continuous use.
Multi-Language Interface English, Spanish, French, Arabic
Password Protected Settings Calibration and data are password protected to avoid accidental deletion
No need for Installation/Commissioning The InfraCal 2 is a portable analyzer, weighing less than 5 lbs. with a 6 in. square foot print, requiring no installation or commissioning, little to no routine maintenance or recalibration.
Minimal Consumables The analysis requires little to no consumables making the cost-per-test minimal.

Specifications

Item Specifications
Analytical Wavelength/Wave Number 9.6 µm, 1045cm-1
Power Requirements 18 volts DC, 3.3 amps, internal battery pack
Power Supply Universal AC/DC provided
Weight 5.8 lb. (2.6 kg) with battery -7.0 lbs (3.2 kg)
Dimensions (17 cm) x 7.8″ (19.8 cm) x 5.2″ (13.2 cm)
Suggested Operating Range 5°C – 40°C
Measurement Range 0 – 20 & 70 – 98% ethanol in gasoline
Analysis Time Less than 30 seconds
Instrument Repeatability 0 – 20%±0.2%, 20 – 98%±1% of ethanol in petrol concentration
Communication Port  USB, RS 232